LIFEPAK 500T AED TRAINING SYSTEM Operating Instructions 23
© 2002-2007 Medtronic Emergency Response Systems, Inc.
Maintenance
Table 8 Setup options troubleshooting
Observation Possible Cause Corrective Action
1 Messages and prompts out of
sequence.
Incorrect setup
options selected.
Review ”Setup Procedures” on
page 16.
2 Messages and prompts
incomplete.
Device damaged. Contact your local Medtronic
representative.
3
CONNECT ELECTRODES
message remains on screen
after training electrodes are
connected to the device.
Instructor did not
press remote control
to begin scenario
sequence.
Damaged remote
control.
IR sensor not
working properly.
Press the ELECTRODE PADS
key once on the remote control
to initiate analysis prompts.
Contact your local Medtronic
representative.
Clean/clear the sensor, if
obstructed. OR
Contact your local Medtronic
representative.
Table 9 Remote control troubleshooting
Observation Possible Cause Corrective Action
1 Device not responding to
remote control key presses.
IR sensor is not
receiving remote
control signal.
Reposition remote control
relative to the AED Trainer.
Remove anything that may be
blocking the IR sensor.
Turn off lights that may be
interfering.
Use remote control cable.
Replace batteries.
Remote control
damaged.
Replace remote control.
Table 10 Training electrodes troubleshooting
Observation Possible Cause Corrective Action
1 Training electrodes do not
adhere to manikin.
Adhesive
contaminated.
Replace training electrodes.
2 Training electrode connector
does not plug into device.
Training electrode
connector damaged.
Replace training electrodes.
Device connector
damaged.
Contact your local Medtronic
representative.