Setup
Configuring Pulse Check
1 In setup mode, press
ANALYZE until the PULSE CHECK screen appears.
2 Press
S to scroll through the following available settings:
NEVER
•AFTER SECOND NSA
AFTER EVERY NSA
•ALWAYS
3 When the desired setting appears, press ANALYZE to save your selection and proceed to the
next setup option, or press
ON/OFF to save your selection and exit setup mode.
Configuring Pulse Prompt
1 In setup mode, press
ANALYZE until the PULSE PROMPT screen appears (on AEDs distributed in
the English language only).
Note: The Pulse Prompt setting is not applicable if Pulse Check is set to
NEVER.
2 Press
S to toggle through the following available settings:
•1CHECK PULSE
•2CHECK PATIENT
3–CHECK FOR SIGNS
3 When the desired setting appears, press ANALYZE to save your selection and proceed to the
next setup option, or press
ON/OFF to save your selection and exit setup mode.
Configuring CPR Time 1 and CPR Time 2
1 In setup mode, press
ANALYZE until the CPR TIME 1 or CPR TIME 2 screen appears (on AEDs
distributed in the English language only).
2 Press
S to scroll through the following available CPR time settings: 0, 15, 30, 45, 60, 90, 120,
or
180 seconds, or 999 (infinite CPR time).
3 When the desired time appears, press
ANALYZE to save your selection and proceed to the next
setup option or press
ON/OFF to save your selection and exit setup mode.
Note:
CPR TIME 0 is not available if AUTO ANALYZE 2 is selected.